To Develop and Validate the Stability Indicating Method for Ezetimibe and
Rosuvastatin Using HPLC Method
Mohd Bilal Sufi; Sakshi Bobade *; Rajlaxmi Deolekar
Department of Pharmaceutics, New Montfort Institute of Pharmacy,
Ashti, Dist. Wardha (M.S) 442202
*Correspondence: sakshibobade2000@gmail.com
DOI: https://doi.org/10.71431/IJRPAS.2026.5718
|
Article
Information
|
|
Abstract
|
|
Review Article
Received: 20/07/2026
Accepted: 21/07/2026
Published:31/07/2026
Keywords
Rosuvastatin,
Ezetimibe, RP-HPLC, Method Validation, Stability-Indicating Method, Forced
Degradation, ICH Q2(R2), Pharmaceutical Analysis.
|
|
A
reverse-phase HPLC (RP-HPLC) method has been developed and validated which is
simple, fast, accurate of Rosuvastatin and Ezetimibe in the pharmaceutical
tablet dose. The separation was done on C18 column (250 × 4.6 mm, 5 µm) with
Methanol acid (60:40, v/v) as mobile phase flow rate of 1.0 mL/min at the
isosbestic wavelength of 233 nm detected using UV detectors. With the
optimized method, well-resolved peaks were obtained for Rosuvastatin around
3.39 min and Ezetimibe around 4.58 min. The method was validated according to
the guidelines outlined by ICH Q2(R2) and was found to be linearly in the
range of 6-30 µg/mL for Rosuvastatin and 3-15 µg/mL for Ezetimibe with
correlation coefficient near unity. The accuracy was found between 98.96 to
100.29% for Rosuvastatin and 98.46–100.74% for Ezetimibe and repeatability,
intraday precision and interday precision showed %RSD values less than 2%
proving the precision of the method. The developed method was determined to
be specific, sensitive, rugged and robust with good limits of detection and
limits of quantitation and validated successfully for the determination of
marketed tablet formulations, giving a high accuracy of 99.81% for
Rosuvastatin, 99.68% of Ezetimibe. Forced degradation studies conducted in
acidic, alkaline, oxidative and photolytic conditions proved that the method
was suitable for identifying degradation products from Rosuvastatin and
Ezetimibe, the latter being relatively more susceptible to degradation than
Rosuvastatin. The developed method of RP-HPLC is simple, economical,
reliable, routine QC method and can be used for stability study of model
compound and simultaneous estimation of Rosuvastatin and Ezetimibe in
combined pharmaceutical dosage formulation.
|
INTRODUCTION:
Hyperlipidemia
(Dyslipidemia) is the abnormal increase, either quantitatively or qualitatively,
of plasma lipoproteins: high total cholesterol, low-density lipoprotein
cholesterol (LDL-C), triglycerides, or a decrease in the amount of high-density
lipoprotein cholesterol (HDL-C) [1-3]. Hyperlipidemia, which is can be divided
into primary (hereditary) and secondary, then classified again as isolated
hypercholesterolemia, isolated hypertriglyceridemia, mixed hyperlipidemia or
low HDL‑C phenotypes [3-4]. Emerging
evidence from population‑based studies suggest that dyslipidemia has become very
common across the world and is one of the most important modifiable risk
factors for atherosclerotic cardiovascular disease, including coronary artery
disease and stroke. [3]
The
epidemiology of dyslipidemia in India has changed dramatically over recent
decades due to social and lifestyle changes associated with rapid urbanization,
sedentary lifestyles, dietary transitions along with large regional variation
[4]. The atherogenic risk attributable to typical Indian dyslipidemia pattern
is also realized through borderline‑high LDL‑C, raised triglycerides and low
HDL‑C [4]. Classical risk factors including obesity, type 2 diabetes mellitus,
hypertension, smoking and metabolic syndrome act synergistically with
dyslipidemia to accelerate atherosclerosis by aggravating development of
endothelial dysfunction and increasing oxidative stress and inflammatory
activation in the arterial wall [3-4].
Atherosclerosis
risk is conferred by the entrapment and oxidation of LDL within the
subendothelial environment, which leads to foam cell formation, plaque
expansion and plaque rupture with clinical sequelae such as myocardial
infarction, ischemic stroke or peripheral arterial disease [3]. Many trials
have established that LDL-C reduction, in particular, but effective lipid-lowering
therapy in general, reduces cardiovascular morbidity and mortality by up to 60%
in both primary and secondary prevention [5-6]. Consequently, currently
existing dyslipidemia management strategies focus on the optimization of
pharmacotherapy by use of agents that affect distinct pathways in cholesterol
homeostasis [5].
To
evaluate a selective cholesterol absorption inhibitor, ezetimibe as an
important adjunctive agent in lipid management with emphasis on patients not
being able to achieve LDL‑C targets with statin therapy alone [5-7].
Chemically, ezetimibe is (3R,4S)‑1‑(4‑fluorophenyl)‑3‑[(S)‑3-(4-fluorophenyl)‐3‐hydroxypropyl]‐4-(4-hydroxyphenyl)azetidine-2‑one,
has the empirical molecular formula C24H21F2NO3
and approximate molecular weight of 409.4 g/mol [8]. The drug is an intestinal
cholesterol absorption inhibitor.
Ezetimibe
is a selective inhibitor of Niemann–Pick C1‑like 1transporter in the intestinal
mucosa, dietary and biliary cholesterol as well as phytosterols are absorbed
from the intestinal lumen by enterocytes in an evolvatively conserved fashion
and mediated primarily by the diffusion of fatty acids across cell membranes or
(in principal) via specific transporters that include NPC1L1. This, in turn,
results in reduced delivery of cholesterol to the liver, up‑regulation of
hepatic LDL receptors and increased clearance of circulating LDL‑C [5, 10].
After oral administration, ezetimibe is quickly absorbed and rapidly conjugated
by phase II metabolism to its pharmacologically active glucuronide conjugate,
which then undergoes enterohepatic recycling that gives it a terminal half‑life
of around 22 hours enabling once‑daily dosing [11-12]. It is predominantly
eliminated by feces, with only a small part excreted by the urine as
glucuronide. [11-12]
On
the therapeutic side, ezetimibe is indicated for primary hypercholesterolemia,
mixed dyslipidemia, homozygous familial high serum cholesterol and homozygous
sitosterolemia. This medicine is usually administered orally as 10 mg
film-coated tablets [7]. Ezetimibe generally has a good safety profile with
mild adverse effects such as soft stool and headache along with increased liver
transaminases, especially in combination with statins. It is also associated with
rare reports of myopathy or rhabdomyolysis[13]. The limiting feature for
ezetimibe therapy is the modest amount by which it lowers LDL‑C when used
alone and the need for careful monitoring when combined with other lipid‑lowering
drugs [5, 7].
Rosuvastatin
is a highly potent statin belonging to the class of synthetic HMG‑CoA reductase
inhibitors with an attractive pharmacokinetic profile. Chemically, rosuvastatin
calcium is (E)‑7‑[4‑(4‑fluorophenyl)−6−isopropyl-2-methyl(methylsulfonyl)amino]pyrimidin-5-y1-3,5-dihydroxyhept‐6‐enoic
acid calcium salt [6]. Its empirical formula is (C22H27FN3O6S)2Ca
and molecular weight 1001.14 g/mol approximately. It is a member of the statin
family and the molecular structure for rosuvastatin. It
is classified as a statin and the detailed chemical structure of rosuvastatin
calcium is provided separately in the dissertation.
Rosuvastatin
is a competitive inhibitor of HMG-CoA reductase, which is the rate-limiting
enzyme involved in hepatic cholesterol biosynthesis; it has been shown to
reduce intrahepatic cholesterol, up-regulate LDL receptors and therefore
increase the clearance of circulating LDL-C [6, 15]. The hepatoselective drug
has a limited ability to invade extrahepatic tissues, and undergoes less
metabolism via cytochrome P450 isoenzymes (as this is the case for other
lipid-lowering drugs), which contribute to its intensive LDL‑C and triglyceride
lowering effects and comparatively lower risk of drug–drug interactions [6-7].
Rosuvastatin is given orally with an effective half‑life of ~19 hours allowing
once‑daily dosing; it is eliminated mainly via fecal excretion [6, 15].
Rosuvastatin
is approved for treatment of primary hypercholesterolemia, mixed dyslipidemia,
and hypertriglyceridemia, as well as to reduce cardiovascular risk in
appropriate high‑risk patients[6, 15]. It is sold under the trade name in
film-coated tablets in strengths of 5 mg, 10 mg, 20 mg and 40 mg. Myalgia,
headache, gastroenteritis and asymptomatic elevation of hepatic transaminases
represent some of the frequent adverse effects; however rare but serious events
like rhabdomyolysis and severe myopathy make clear that patient selection and
monitoring is essential. The limitations of rosuvastatin therapy include statin
intolerance in some [6, 15].
Combining
ezetimibe with rosuvastatin is scientifically rational because the two agents
act on distinct yet complementary pathways of cholesterol homeostasis [5, 10].
Because the cholesterol-lowering effect of statins occurs by two mechanisms as
inhibition of both endogenous production and receptor-mediated clearance of
LDL-C; ezetimibe reduces intestinal absorption of exogenous and biliary
cholesterol, thereby providing dual inhibition of cholesterol influx and
production. [5, 10] Clinical studies have
demonstrated that co‑administration of ezetimibe plus statins, including
rosuvastatin, yields larger reductions in LDL‑C and non‑HDL‑C compared with
statin monotherapy — together with beneficial effects on triglycerides and HDL‑C.
[5-6]
This
dual-action therapy approach has specific utility in patients with high
baseline LDL‑C, familial hypercholesterolemia, or statin intolerance at higher
doses and allows attaining target lipid levels with lower statin dosages [5]. Ezetimibe–rosuvastatin
functional combinations improve surrogate markers of atherosclerosis and also
prevent the incidence of major cardiovascular events with additional benefit in
high-risk populations when compared to statin monotherapy, improving lipid
control through additive effects on low-density lipoprotein (LDL) particle size
and non-HDL cholesterol levels. Consequently, fixed-dose combination
formulations of ezetimibe and rosuvastatin have become one of the mainstay
therapeutic options; strong analytical procedures are necessary for routine
quality control and efficacy studies [1-2].
For many chemically
diverse drugs, reversed-phase HPLC (RP‑HPLC), with a non‑polar stationary
phase, such as C18 and a relatively polar mobile phase containing aqueous
mobile phase and organic modifiers, is the best available method for separating
and quantifying components, in particular because of its robustness and
suitability for both the aqueous and organic solvents common to many drug
formulations, as well as for compatibility with UV detection.
[18-19]
The selection of stationary phase, mobile phase composition and pH, column flow
rate, column temperature, and detection wavelengths are very important to
achieve best resolution, peak symmetry, and optimum reproducibility while
simultaneously estimating multiple drugs in complex matrices in HPLC systems.
RP-HPLC is very well
suited for pharmaceutical analysis of fixed-dose combinations including the
combination of ezetimibe and rosuvastatin, which contains aromatic chromophores
and moderate lipophilicity of both drugs [19, 21]. RP-HPLC is well suited to
the pharmaceutical analysis of fixed dose combinations such as ezetimibe and
rosuvastatin, which contains aromatic chromophores and a moderate lipophilicity
of both drugs.
A stability indicating
analytical method is defined, in line with regulatory and ICH guidelines, as a
validated analytical method that quantifies the pharmaceutical product or drug
substance accurately and precisely and demonstrates for the assay profile that
the product is free from process impurities, excipients and degradation
products, both normal storage conditions and stress induced ones. [18,23] These
methods are extremely essential for establishing the intrinsic stability of a
pharmaceutical product, to support shelf life development, and for assuring
quality, safety and efficacy throughout the product lifecycle. [24-25] For
combination products, the method should be able to resolve and quantify each
active pharmaceutical ingredient (API), their respective degradants and
additionally the combination product itself, which can prove to be challenging
during method development and optimization. [18, 23] For a
combination product, an analytical method should be able to separate and
quantify each active pharmaceutical ingredient, their respective degraded
forms, and also the combination product itself, which can be challenging in the
development and optimization of the method. [1, 26]
Most of the essential
validation parameters recommended in the revised ICH Q2(R2) Guideline for the
Validation of Analytical Procedures include the following for assay methods
used for quantitative chromatographic analysis such as specificity,
selectivity, linearity, accuracy, precision, range, limit of detection (LOD),
limit of quantification (LOQ), robustness, ruggedness (where applicable), and
system suitability testing.
Specificity and
selectivity describe the ability of the method to measure the analyte
unequivocally in the presence of other components such as impurities,
excipients, and degradants, which is particularly important for
stability‑indicating assays.[23, 27] Linearity refers to the proportional
relationship between analyte concentration and detector response over a defined
range, while accuracy expresses the closeness of measured values to the true
value, often assessed via recovery studies. [21, 27] Precision encompasses
repeatability and intermediate precision, evaluating the degree of agreement
between replicate measurements under the same and varying conditions,
respectively.[27, 30-31]
LOD and LOQ represent the
smallest amounts of analyte that can be reliably detected and quantified, which
are critical for impurity and degradation product profiling.[23, 27] Robustness
reflects the method’s capacity to remain unaffected by small deliberate
variations in parameters such as mobile phase composition, pH, flow rate, and
column temperature, thereby indicating its reliability during routine use. [22,
27] Ruggedness, when assessed, extends this concept to inter‑laboratory
changes, different analysts, or instruments. [25, 27] System suitability tests,
including parameters such as theoretical plates, tailing factor, resolution,
and repeatability of peak response, are performed prior to analysis to verify
that the chromatographic system is functioning properly.[20, 27]
Several
chromatographic methods for the simultaneous determination of ezetimibe and
rosuvastatin in bulk drug and combination tablet dosage forms have been
reported, a few of which contained stability-indicating element(s), however,
most of the above mentioned methods involve relatively complex mobile phases,
longer run time, limited evaluation of forced degradation conditions and
incompleteness of validation meeting current ICH Q2(R2) recommendation for
analytical procedure. [1-2,26] Further, the increased use of
ezetimibe–rosuvastatin fixed-dose products in human clinical practice requires
simple, rapid, accurate, precise, sensitive, robust and economical method to be
adopted in routine quality control laboratories. [5-6] The present study is of
immense importance since it aims to develop and validate a novel, stability
indicating RP‑HPLC method for the simultaneous estimation of ezetimibe and
rosuvastatin in pharmaceutical dosage form, which not only meets current
regulatory requirement and analytical science but will also offer a well
captured stability indication data for the assay of ezetimibe and rosuvastatin
[27, 29] simultaneously in longer period
of time under the ICH recommended stress and storage conditions which will
facilitate formulation development, shelf life assignment and submission to
regulatory authority.
MATERIALS & METHODS
Rosuvastatin
and Ezetimibe active pharmaceutical ingredients (APIs) were procured from
Swapnroop Drugs and Pharmaceuticals, Aurangabad, India. The marketed formulation,
Novastat EZ tablets, was obtained from Lupin Ltd. HPLC-grade methanol and 0.1 N
citric acid were purchased from Merck Ltd., India, while purified water was
obtained from Research Lab Fine Chem Industries. All chemicals and reagents
used in the study were of analytical or HPLC grade and were used without
further purification.
Determination
of Absorption Spectra
The
Rosuvastatin and Ezetimibe stock solutions were separately dissolved in
methanol to make the concentration of 1000 µg/mL from accurately weighted
quantities of each drug. Working standard drug solutions of 10µg/mL were
prepared as appropriate dilutions of the drug solutions. The UV absorption spectra were recorded over the
wavelength range of 200–400 nm using methanol as the blank. The individual
absorption maxima (λmax) of Rosuvastatin and Ezetimibe were determined from the
spectra, and the overlain spectra were used to identify the common isosbestic
wavelength, which was selected for subsequent RP-HPLC analysis.
HPLC Instrumentation and Chromatographic Conditions
Chromatographic analysis was performed using an Agilent 1100 Gradient
High-Performance Liquid Chromatography (HPLC) system equipped with a G1310A
ISOPUMP, a Photodiode Array (PWD) detector, and ChemStation software for
instrument control, data acquisition, and analysis. Separation was achieved on
an Agilent RP-C18 column (4.6 × 250 mm, 5.0 µm particle size) using a mobile
phase consisting of methanol and 0.1 N citric acid (60:40, v/v). The mobile
phase was delivered at a flow rate of 1.0 mL/min, and analytes were detected at
a wavelength of 233 nm. The column temperature was maintained at 25°C, and the
injection volume was 20 µL. The G1310A ISOPUMP is a high-pressure reciprocating
pump with a maximum operating pressure of 400 bar, a flow rate range of
0.001–5.0 mL/min, pressure display accuracy of ±5%, four-solvent capability,
and a programmable mixing ratio of 0–100%, providing precise and reproducible
mobile phase delivery throughout the chromatographic analysis.
Method
Development and Optimization
Chromatographic conditions were optimized by evaluating different mobile
phase compositions consisting of methanol with 0.1% acetic acid and citric acid
in various proportions. A variety of
combinations were subjected to flow rates suitable for the application (0.7 to
1.0 mL/min) to obtain good peak resolution, retention time, peak shape, and
theoretical plate count. The optimized mobile phase of Methanol : Citric acid
(60:40, v/v) was chosen based on the chromatographic performance. The optimized
method gave well-resolved and symmetrical peaks for both analytes having
satisfactory system suitability parameters.
Table 01 Method Development Trials for Optimization of RP-HPLC
Conditions
|
Trial No.
|
Mobile Phase
Composition (v/v)
|
Flow Rate
(mL/min)
|
|
Trial 01
|
Methanol : 0.1% Acetic Acid (90:10)
|
0.7
|
|
Trial 02
|
Methanol : 0.1% Acetic Acid (85:15)
|
0.7
|
|
Trial 03
|
Methanol : 0.1% Acetic Acid (80:20)
|
0.7
|
|
Trial 04
|
Methanol : 0.1% Acetic Acid (75:25)
|
0.7
|
|
Trial 05
|
Methanol : 0.1% Acetic Acid (85:15)
|
1.0
|
|
Trial 06
|
Methanol : 0.1% Acetic Acid (88:12)
|
1.0
|
|
Trial 07
|
Methanol : 0.1% Citric Acid (80:20)
|
0.7
|
|
Trial 08
|
Methanol : 0.1% Citric Acid (60:40)
|
1.0
|
|
Trial 09
|
Methanol : 0.1% Citric Acid (55:45)
|
1.0
|
|
Trial 10
|
Methanol : 0.1% Citric Acid (55:45)
|
1.0
|
|
Trial 11
|
Methanol : 0.1 N Citric Acid (60:40)
|
1.0
|
Preparation
of Standard Stock Solution
Accurately
weighed quantities of Rosuvastatin (30 mg) and Ezetimibe (15 mg) were
transferred into separate 100 mL volumetric flasks. About 70 mL of methanol was
then added to each solution and sonicated for 15 minutes to allow them to
completely dissolve. The concentrations of Rosuvastatin and Ezetimibe were then
adjusted to be 300 µg/mL and 150 µg/mL in methanol to form stock solutions.
Preparation
of Working Standard Solutions (Linearity)
Standard
stock solutions were suitably diluted with the optimized mobile phase to make
the calibration standards. Rosuvastatin working solutions were diluted in the
range of 6–30 µg/mL, and Ezetimibe solutions in the range of 3–15 µg/mL. All
solutions were sonicated for the removal of dissolved gases, and filtered
through a 0.45 µm membrane filter before the injection.
Accuracy
Standard
addition method was used to assess the developed RP-HPLC method of the accuracy
up to 80, 100 and 120% of the target concentration. Previously analysed tablet sample solutions were
spiked with known quantities of Rosuvastatin and Ezetimibe standard solutions
and analyzed under the optimized chromatographic conditions.
To determine the accuracy of the analytical method, the percentage recovery was
calculated for each level.
Repeatability
The
repeatability of the method was determined by processing six replicate
injections of a mixed standard solution, containing Rosuvastatin (24 µg/mL) and
Ezetimibe (12 µg/mL), under similar conditions. The repeatability of the
proposed RP-HPLC method was evaluated by calculating the mean, standard
deviation and percentage relative standard deviation (%RSD) from the peaks
areas obtained from the replicate injections.
Intraday
Precision
Intra-day
precision was evaluated by analyzing a mixed standard solution with three
different concentrations of Rosuvastatin (12, 18 and 24 µg/mL) and three
different concentrations of Ezetimibe (6, 9 and 12 µg/mL) at various timepoints
within the same day. All concentrations were analyzed in duplicate and
conditions were the same on the chromatograph for all. The precision was
reported as the percent relative standard deviation (%RSD) of the peak areas.
Interday
Precision
Interday
precision was assessed by analyzing freshly prepared mixed standard solutions
containing 12, 18, and 24 µg/mL of Rosuvastatin and 6, 9, and 12 µg/mL of
Ezetimibe on three consecutive days using the same chromatographic conditions.
The consistency of the analytical method was evaluated by calculating the mean
peak area, standard deviation, and percentage relative standard deviation
(%RSD).
Limit
of Detection (LOD) and Limit of Quantification (LOQ)
The
developed RP-HPLC method was evaluated for its sensitivity and the limits of
detection (LOD) and limits of quantification (LOQ) were determined as per ICH
Q2(R1). The parameters can be computed from the standard deviation of the
response (σ) and the slope (S) of the calibration curve as follows:
The
calculated LOD and LOQ values showed that the developed analytical method was
sensitive enough for the simultaneous estimation of Rosuvastatin and Ezetimibe.
Specificity
The
developed HPLC method was validated for specificity by analyzing the blank and
standard solutions as well as sample solutions under the optimized HPLC
conditions. The chromatograms were checked for the presence of any excipient
peak, solvent peak or impurity peak at the retention time of Rosuvastatin and
Ezetimibe. The two peaks, namely the interfering peaks, were not seen in the
background, indicating that the background was contained and not interfering
with the method developed.
Ruggedness
The
developed method was evaluated for ruggedness by analyzing the samples by the
same method, but on different days and different analysts. A standard solution
of 18µg/mL for Rosuvastatin and 9µg/mL for Ezetimibe was analyzed and the
obtained results were compared by the following parameters: peak area,
percentage label claim. The %RSDs were small, which indicated the ruggedness of
the method.
Robustness
To
test the robustness of the RP-HPLC method the following deliberate changes were
introduced in the chromatographic conditions: mobile phase composition ±1% and
wavelength of detection (243 and 245 nm), with all other experimental
conditions remaining the same. Under these modified conditions, standard
solutions of 18 µg/mL of Rosuvastatin in 9 µg/mL of Ezetimibe were analyzed.
Areas and the %RSD values generated were examined to assess the effect of these
minor variations on the method.
Assay
of Marketed Tablet Formulation
Twenty marketed tablets of Novastat EZ tablets containing Rosuvastatin
and Ezetimibe were accurately weighed and finely powdered.
Both powders contained the same mass on a label as each other and the same mass
was transferred into a volumetric flask and extracted with methanol, sonicated
for 15 minutes after which each was filtered through a 0.45 µm membrane filter.
The optimized mobile phase was used to prepare suitable dilutions in order to
obtain the final concentration of rosuvastatin and ezetimibe 18 µg/mL and 9
µg/mL, respectively. The optimal chromatography conditions were used to analyze
the prepared solution and the percentage label claim was determined.
Forced
Degradation Studies
Acid
Degradation
Acid
degradation was performed by adding appropriate amount of 0.1 N HCl to the
standard solution mixture. The solution was left to stand for 2 hours and then
for 24 hours and before the chromatographic analysis the solution was
neutralized and diluted with the mobile phase appropriately.
Alkaline
Degradation
The
mixed standard solution was treated with 0.1 N sodium hydroxide to degrade it
in an alkaline condition and then kept for 2 hours and 24 hours at room
temperature. The solution was subjected to the mobile phase after
neutralization and filtered with 0.45 µm membrane filter and then analyzed by
RP-HPLC.
Oxidative Degradation
Oxidative
degradation test was carried out by reacting the mixed standard solution with
3% hydrogen peroxide (H₂O₂). The solution was stabbed for 2 h and 24 h under
ambient laboratory conditions and analyzed by analysis of chromatography to get
the solution. The areas of the sample and the standard were compared to find
the percentage degradation.
Photolytic
Degradation
To
simulate the photo-degradation, the mixture standard solution was subjected to
UV radiation for the mentioned exposure period. The solution exposed was
analysed by the optimized RP-HPLC method and the percentage of degradation was
calculated by comparing the peak areas of the exposed sample with the peak
areas of the untreated sample.
RESULTS AND DISCUSSION:
Determination
of Absorption Spectrum and Iso-absorptive Point
The UV absorption spectra of Rosuvastatin and Ezetimibe were recorded
individually in methanol over the wavelength range of 200–400 nm to determine
their respective absorption maxima and identify a suitable analytical
wavelength for simultaneous estimation. The maximum
absorbance of rosuvastatin and ezetimibe were found to be 236 nm and 227 nm
respectively. The two drugs had an isosbestic curve point at 233nm between the
overlain spectra of both drugs, indicating that these two analytes have the
same absorptivity at this wavelength. Thus, the detection wavelength of 233 nm
was chosen for further RP-HPLC analysis, which allowed the determination with
proper sensitivity and simultaneous determination of both drugs without
significant spectral interferences.
Fig. 01 UV absorption spectrum of Rosuvastatin (10 µg/mL) in
methanol
Fig. 02 UV absorption spectrum of Ezetimibe (10 µg/mL) in
methanol
Fig. 03 Isosbestic point of Rosuvastatin and Ezetimibe at 233 nm
Mobile
Phase Optimization
Optimization of the chromatographic conditions was carried out through
systematic evaluation of different mobile phase compositions consisting of
methanol combined with 0.1% acetic acid and 0.1% citric acid at various
proportions and flow rates. During the first
trials, a methanol–acetic acid system was employed, resulting in wide peaks,
poor peak shape and poor resolution of Rosuvastatin and Ezetimibe. Optimization
with methanol–citric acid mixtures gave better chromatographic performance. The
optimized chromatographic condition were found to be with Methanol : Citric
acid (60:40, v/v) with flow rate of 1.0 mL/min where retention time of
Rosuvastatin and Ezetimibe were around 3.39 and 4.58 min, with their sharp peak
shape and symmetrical peak with good resolution. The optimized chromatogram
showed good theoretical plates, acceptable peak symmetry, and complete baseline
resolution required for the simultaneous estimation and further validation
studies and the mobile phase was found to be suitable.
Fig. 04 Chromatogram of Trial-11 optimized mobile phase
condition
Table 02 Chromatographic performance data of mobile phase
trials
|
Trial
|
Mobile
Phase Composition
|
Peak
|
Retention
Time (min)
|
Area
(mAU·s)
|
Height
(mAU)
|
Symmetry
|
Width
(min)
|
Plates
|
|
01
|
MeOH 90% + 0.1% Acetic Acid 10% (Flow 0.7
mL/min)
|
Peak 1
|
4.524
|
210.91
|
17.99
|
5.08
|
0.2103
|
2567
|
|
Peak 2
|
4.603
|
153.65
|
18.48
|
0.44
|
–
|
–
|
|
Peak 3
|
4.724
|
68.66
|
15.97
|
0.34
|
–
|
–
|
|
Peak 4
|
4.843
|
158.78
|
20.91
|
0.65
|
0.1262
|
8151
|
|
Peak 5
|
6.349
|
77.86
|
3.35
|
0.51
|
0.3622
|
1701
|
|
Peak 6
|
7.514
|
448.98
|
28.09
|
0.99
|
0.2344
|
5694
|
|
02
|
MeOH 85% + 0.1% Acetic Acid 15% (Flow 0.7
mL/min)
|
Peak 1
|
4.485
|
194.16
|
15.56
|
2.00
|
0.2151
|
2410
|
|
Peak 2
|
4.647
|
242.79
|
16.68
|
0.42
|
–
|
–
|
|
Peak 3
|
4.932
|
209.60
|
22.91
|
0.73
|
0.1456
|
6364
|
|
Peak 4
|
6.847
|
80.57
|
2.58
|
0.43
|
0.4500
|
1283
|
|
Peak 5
|
10.770
|
482.20
|
19.07
|
1.03
|
0.3600
|
4962
|
|
03
|
MeOH 80% + 0.1% Acetic Acid 20% (Flow 0.7
mL/min)
|
Peak 1
|
4.639
|
22.96
|
2.81
|
0.49
|
0.1518
|
5174
|
|
Peak 2
|
4.795
|
11.61
|
2.77
|
1.18
|
0.0818
|
19071
|
|
Peak 3
|
5.054
|
341.20
|
36.81
|
0.73
|
0.1387
|
7351
|
|
Peak 4
|
7.561
|
50.37
|
1.66
|
0.69
|
0.5867
|
919
|
|
04
|
MeOH 75% + 0.1% Acetic Acid 25% (Flow 0.7
mL/min)
|
Peak 1
|
4.685
|
54.08
|
3.84
|
1.62
|
0.2622
|
1767
|
|
Peak 2
|
4.865
|
31.45
|
4.70
|
1.07
|
0.1331
|
7398
|
|
Peak 3
|
5.255
|
428.27
|
36.50
|
1.02
|
0.1578
|
6143
|
|
05
|
MeOH 85% + 0.1% Acetic Acid 15% (Flow 1.0
mL/min)
|
Peak 1
|
5.022
|
428.21
|
47.39
|
0.62
|
0.1343
|
7745
|
|
Peak 2
|
11.124
|
391.56
|
22.92
|
0.66
|
0.2633
|
9882
|
|
06
|
MeOH 88% + 0.1% Acetic Acid 12% (Flow 1.0
mL/min)
|
Peak 1
|
3.491
|
277.18
|
34.58
|
1.28
|
0.1300
|
3992
|
|
Peak 2
|
6.586
|
259.72
|
25.02
|
0.67
|
0.1617
|
9186
|
|
07
|
MeOH 80% + 0.1% Citric Acid 20% (Flow 0.7
mL/min)
|
Peak 1
|
9.556
|
965.84
|
62.96
|
0.61
|
0.2367
|
9029
|
|
Peak 2
|
12.854
|
1226.80
|
61.50
|
0.62
|
0.3114
|
9438
|
|
08
|
MeOH 60% + 0.1% Citric Acid 40% (Flow 1.0
mL/min)
|
Peak 1
|
4.004
|
589.05
|
68.99
|
0.58
|
0.1290
|
5330
|
|
Peak 2
|
4.860
|
250.46
|
25.01
|
0.56
|
0.1489
|
5906
|
|
09
|
MeOH 55% + 0.1% Citric Acid 45% (Flow 1.0
mL/min)
|
Peak 1
|
4.394
|
591.88
|
60.32
|
0.60
|
0.1444
|
5129
|
|
Peak 2
|
5.757
|
250.86
|
21.00
|
0.59
|
0.1756
|
5954
|
|
10
|
MeOH 55% + 0.1% Citric Acid 45% (Flow 1.0
mL/min)
|
Peak 1
|
4.425
|
603.48
|
59.80
|
0.59
|
0.1477
|
4976
|
|
Peak 2
|
5.814
|
248.96
|
20.49
|
0.60
|
0.1800
|
5776
|
|
11
|
MeOH 60% + Citric Acid 40% (Flow 1.0 mL/min)
|
Rosuvastatin
|
3.385
|
1338.41
|
140.44
|
0.95
|
0.1433
|
3088
|
|
Ezetimibe
|
4.582
|
684.12
|
58.11
|
0.99
|
0.1797
|
3601
|
Linearity
The
linearity of developed RP-HPLC method was investigated by analyzing standard
solution of Rosuvastatin in the range of 6–30 µg/mL and Ezetimibe in the range
of 3–15 µg/mL. For both drugs, a linear relationship between peak area and
concentration was determined over the given concentration ranges. Rosuvastatin
showed a regression equation of y = 113.1175x + 73.687 with a correlation
coefficient (R² = 0.9998), and Ezetimibe showed an excellent linearity with a
correlation coefficient very close to unity. The obtained low standard
deviation and percentage relative standard deviation (%RSD) showed good
repeatability in the responses of detector at all the concentrations. These
results verified the method developed is useful for the intended analytical
concentration range and meets the ICH guidelines for the linearity.
Table 03 Linearity Data for Rosuvastatin (ROSU)
|
Concentration
(µg/mL)
|
Mean
Area
|
SD
|
%RSD
|
|
6
|
745.58
|
6.79
|
0.91
|
|
12
|
1438.18
|
0.32
|
0.02
|
|
18
|
2124.04
|
3.04
|
0.14
|
|
24
|
2766.03
|
0.20
|
0.01
|
|
30
|
3475.18
|
0.89
|
0.03
|
Table 04 Linearity Data for Ezetimibe (EZET)
|
Concentration
(µg/mL)
|
Mean
Area
|
SD
|
%RSD
|
|
3
|
343.85
|
5.08
|
1.48
|
|
6
|
674.76
|
0.89
|
0.13
|
|
9
|
985.61
|
5.65
|
0.57
|
|
12
|
1306.03
|
4.15
|
0.32
|
|
15
|
1642.17
|
0.25
|
0.01
|
Fig. 04 Calibration curve of Rosuvastatin
Fig. 05 Calibration curve of Ezetimibe
Accuracy
The
developed method was checked for the accuracy by standard addition method at
80%, 100% and 120% concentration levels. The mean percentage recoveries of
Rosuvastatin ranged between 98.96% to 100.29% whilst the Ezetimibe mean percentage
recoveries ranged between 98.46% to 100.74%. The RSD percentage value for both
the drugs was below 2%, which also indicates excellent reproducibility of the
recovery results. The recovery values were observed to be approaching 100%,
indicating that the developed analytical method was effective in quantifying
both the analytes without getting interfered by the formulation excipients. The
results indicated that proposed RP-HPLC method was highly accurate and useful
for routine pharmaceutical analysis.
Table 05 Summary of Accuracy Studies of Rosuvastatin
(ROSU)
|
Sample
(% level)
|
Conc.
of marketed formulation (µg/mL)
|
Conc.
of API added (µg/mL)
|
Absorbance
(Area)
|
Amount
Recovered (µg/mL)
|
%
Recovery
|
Mean
|
SD
|
%RSD
|
|
80%
|
12
|
9.6
|
2503.68
|
9.49
|
98.81
|
98.96
|
0.22
|
0.22
|
|
2510.00
|
9.51
|
99.11
|
|
100%
|
12
|
2792.52
|
12.04
|
100.33
|
100.29
|
0.06
|
0.05
|
|
2791.47
|
12.03
|
100.25
|
|
120%
|
14.4
|
3053.78
|
14.35
|
99.65
|
99.44
|
0.30
|
0.30
|
|
3050.00
|
14.29
|
99.23
|
Table 06 Summary of Accuracy Studies of Ezetimibe (EZET)
|
Sample
(% level)
|
Conc.
of marketed formulation (µg/mL)
|
Conc.
of API added (µg/mL)
|
Absorbance
(Area)
|
Amount
Recovered (µg/mL)
|
%
Recovery
|
Mean
|
SD
|
%RSD
|
|
80%
|
6
|
4.8
|
1173.28
|
4.74
|
98.72
|
98.46
|
0.37
|
0.38
|
|
1170.57
|
4.71
|
98.19
|
|
100%
|
6
|
1307.60
|
5.99
|
99.86
|
99.60
|
0.36
|
0.37
|
|
1304.29
|
5.96
|
99.34
|
|
120%
|
7.2
|
1445.47
|
7.28
|
101.08
|
100.74
|
0.48
|
0.48
|
|
1440.24
|
7.23
|
100.40
|
Intraday
Precision
Intraday
precision of the developed RP-HPLC method was evaluated by analyzing three
different concentrations of Rosuvastatin (12, 18, and 24 µg/mL) and Ezetimibe
(6, 9, and 12 µg/mL) multiple times within the same day under identical
chromatographic conditions. The assay values obtained in the process were found
to be close to the nominal value, with the %RSD of the values in the range of
0.05-0.27% for Rosuvastatin and 0.00-0.38% for Ezetimibe. The %RSD data were
indicative of high repeatability and consistency between replicate analyses
performed on the same day. As a result these results were found to be in
accordance with the developed RP-HPLC method having high intraday precision and
also fulfilling the acceptance criteria fixed by ICH guidelines.
Interday
Precision
Interday
precision was assessed by analyzing freshly prepared standard solutions of
Rosuvastatin (12, 18, and 24 µg/mL) and Ezetimibe (6, 9, and 12 µg/mL) on
different days using the same chromatographic conditions. The values obtained
at the various time points of the study were comparable indicating the time
stability of the developed method. . The %RSD values for Ezetimibe ranged from
0.12% to 0.78%, while Rosuvastatin exhibited acceptable precision across all
concentration levels, confirming reliable analytical performance between
different days. The results obtained confirmed the developed HPLC method
yielded acceptable intermediate precision for the simultaneous RP-HPLC analysis
of Rosuvastatin and Ezetimibe in tablet formulation, which can be used for
routine quality control analysis.
Table 07 Intraday Precision for Rosuvastatin (ROSUVA)
|
Concentration
(µg/mL)
|
Mean
Area
|
Amount
Found (µg/mL)
|
%
Amount Found
|
SD
|
%RSD
|
|
12
|
1439.26
|
12.07
|
100.62
|
0.66
|
0.05
|
|
18
|
2133.50
|
18.21
|
101.18
|
5.68
|
0.27
|
|
24
|
2768.89
|
23.83
|
99.29
|
4.37
|
0.16
|
Table 08 Intraday Precision for Ezetimibe (EZET)
|
Concentration
(µg/mL)
|
Mean
Area
|
Amount
Found (µg/mL)
|
% Amount
Found
|
SD
|
%RSD
|
|
6
|
674.10
|
6.08
|
101.37
|
0.03
|
0.00
|
|
9
|
984.02
|
8.97
|
99.70
|
1.83
|
0.19
|
|
12
|
1308.03
|
12.00
|
99.96
|
4.91
|
0.38
|
Table 09 Interday Precision for Rosuvastatin (ROSUVA)
|
Concentration
(µg/mL)
|
Mean
Area
|
Amount
Found (µg/mL)
|
%
Amount Found
|
SD
|
%RSD
|
|
12
|
1439.21
|
12.07
|
100.61
|
45.25
|
3.14
|
|
18
|
2089.73
|
17.83
|
99.03
|
8.82
|
0.42
|
|
24
|
2802.30
|
24.13
|
100.52
|
0.57
|
0.02
|
Table 10 Interday Precision for Ezetimibe (EZET)
|
Concentration
(µg/mL)
|
Mean
Area
|
Amount
Found (µg/mL)
|
%
Amount Found
|
SD
|
%RSD
|
|
6
|
674.45
|
6.09
|
101.42
|
5.24
|
0.78
|
|
9
|
990.54
|
9.03
|
100.38
|
6.08
|
0.61
|
|
12
|
1302.32
|
11.94
|
99.52
|
1.63
|
0.12
|
LOD
and LOQ:
The
sensitivity of the developed RP-HPLC method was determined by calculating the
Limit of Detection (LOD) and Limit of Quantification (LOQ). The obtained values
proved that developed chromatographic method has excellent analytical
sensitivity, which allows detecting and accurate quantifying of very low
concentration levels of Rosuvastatin and Ezetimibe. The results of the proposed
method with low detection limits and low quantitative limits revealed that the
method has great potential applications in quality control routine
applications, stability studies and can be used for the determination of trace
level in pharmaceutical formulations.
Repeatability
Repeatability
was assessed after six injections of standard solutions were made, with the
same chromatographic conditions. The mean peak area for Rosuvastatin was 98.61%
assay at 24 µg/mL, whereas that of Ezetimibe was 99.74% assay at 12 µg/mL with
a fortuitously low %RSD of 0.02%. The low overall variability of each injection
proved the precision of the instrument and the uniformity of the developed
chromatographic method. These findings demonstrated the good repeatability of
the method for general quality control.
Table 11 Repeatability for Rosuvastatin and Ezetimibe
|
Drug
|
Concentration
(µg/mL)
|
Mean
Area
|
Amount
Found (µg/mL)
|
%
Amount Found
|
SD
|
%RSD
|
|
Rosuvastatin
|
24
|
2750.41
|
23.67
|
98.61
|
22.86
|
0.83
|
|
Ezetimibe
|
12
|
1305.11
|
11.97
|
99.74
|
0.30
|
0.02
|
Specificity
Specificity
studies were performed, and the developed RP-HPLC method was able to
selectively estimate Rosuvastatin and Ezetimibe in the presence of formulation
excipients, without any chromatographic interference between the two and any
excipients. For both drugs, well-resolved peaks were generated with good peak
purity, and consistent peak retention times. The assay of Rosuvastatin was an
average of 100.75%, and Ezetimibe, 99.59% with very low %RSD values of 0.065%
and 0.056%, respectively. These results indicated that there was no
interference of the common excipients and other formulation components during
the determination of analytes and the developed analytical method was
considered to be specific.
Table 12 Specificity for Rosuvastatin and Ezetimibe
|
Drug
|
Concentration
(µg/mL)
|
Mean
Area
|
Amount
Found (µg/mL)
|
% Label
Claim
|
SD
|
%RSD
|
|
Rosuvastatin
|
18
|
2124.75
|
18.14
|
100.75
|
0.012
|
0.065
|
|
Ezetimibe
|
9
|
982.93
|
8.96
|
99.59
|
0.005
|
0.056
|
Ruggedness
Ruggedness
was checked by comparing the same sample tested under varying analytical
parameters with different analysts and different operating days. Rosuvastatin
had an average assay of 100.79% whereas Ezetimibe exhibited 99.57%,
substantially low %RSD of 0.016% and 0.085% respectively. The method being
developed was also significantly rugged and reproducible as minimal difference
was seen between the different operating conditions.
Table 13 Ruggedness Study for Rosuvastatin and Ezetimibe
|
Drug
|
Conc
(µg/mL)
|
Amount
Found (Mean, µg/mL)
|
% Label
Claim (Mean)
|
SD
|
%RSD
|
|
Rosuvastatin
|
18.00
|
18.14
|
100.79
|
0.003
|
0.016
|
|
Ezetimibe
|
9.00
|
8.96
|
99.57
|
0.008
|
0.085
|
Robustness
The
robustness of the developed analytical method was investigated by deliberately
introducing minor variations in chromatographic parameters, including mobile
phase composition and detection wavelength. All deliberate changes did not
result in any noticeable changes in retention time, peak shape, peak area, or
chromatographic resolution. The %RSD for each value obtained were less than 2%,
showing that variations in the experimental conditions caused little effect on
the analytical performance. Hence, the developed method RP-HPLC proved to be
sufficiently robust and reliable method and hence could be used in regular
pharmaceutical analysis.
Table 14 Robustness Study for Rosuvastatin and Ezetimibe
|
Parameter
|
Drug
|
Condition
|
Conc
(µg/mL)
|
Area
|
Mean
|
SD
|
%RSD
|
|
Mobile Phase Change
|
Ezetimibe
|
39:61 MeOH
|
9
|
1029.43
|
1027.20
|
3.10
|
0.30
|
|
1025.05
|
|
41:59 MeOH
|
1024.64
|
1024.42
|
0.32
|
0.03
|
|
1024.19
|
|
Rosuvastatin
|
39:61 MeOH
|
18
|
2150.00
|
2149.20
|
0.88
|
0.04
|
|
2148.40
|
|
41:59 MeOH
|
2142.49
|
2142.32
|
0.17
|
0.01
|
|
2142.15
|
|
Wavelength Change
|
Ezetimibe
|
243 nm
|
9
|
1008.53
|
1020.90
|
17.45
|
1.71
|
|
1033.21
|
|
245 nm
|
314.01
|
315.73
|
2.43
|
0.77
|
|
317.44
|
|
Rosuvastatin
|
243 nm
|
18
|
2161.89
|
2158.50
|
3.39
|
0.16
|
|
2155.11
|
|
245 nm
|
3016.09
|
3015.79
|
0.30
|
0.01
|
|
3015.49
|
Assay
of Tablet Formulation
The
RP-HPLC method was validated and successfully applied to determine the
quantitative content of Rosuvastatin and Ezetimibe in the Novastat EZ tablets.
Rosuvastatin had a mean assay value of 100.81% of the label claim, while
Ezetimibe showed 99.68% of the label claim. This was reflected in the low SD
and %RSD values that indicated high assay repeatability as well as accuracy.
The developed procedure proved to be appropriate for quality control
application of the combined formulation of tablet dosage forms of Rosuvastatin
and Ezetimibe on daily basis.
Table 15 Assay of Rosuvastatin and Ezetimibe (% Label
Claim)
|
Drug
|
Conc
(µg/mL)
|
Area I
|
Area II
|
Mean
Area
|
Amount
Found (Mean, µg/mL)
|
% Label
Claim (Mean)
|
SD
|
%RSD
|
|
Rosuvastatin
|
18.00
|
2105.64
|
2105.54
|
2105.59
|
17.97
|
99.81
|
0.004
|
0.004
|
|
Ezetimibe
|
9.00
|
982.63
|
985.04
|
983.84
|
8.97
|
99.68
|
0.178
|
0.178
|
Fig 06: Chromatogram of Marketed
formulation
Force degradation:
Acid
Degradation
The
developed RPHPLC method was evaluated in acidic stress conditions to check the
stability of Rosuvastatin and Ezetimibe in the acidic environment. The rate of
degradation of Ezetimibe and Rosuvastatin by acid hydrolysis was 3.83% and
6.79%, respectively, after 2 h, suggesting that Rosuvastatin is relatively
highly susceptible to hydrolysis by acid. With a longer exposure time of 24 hours
degradation rose to 7.09 % for Ezetimibe and 9.62 % for Rosuvastatin. The
developed RP-HPLC method is also highly specific and showed high stability
indicating ability, as the peaks due to the formed degradation products were
well separated from the analyte peaks under the acidic stress conditions
without any interference.
Alkaline
(Basic) Degradation
Alkaline
stability of the both drugs was studied by exposing the samples to alkaline
hydrolysis. Ezetimibe exhibited 3.97% degradation and Rosuvastatin exhibited
6.85% degradation after 2 hours indicating Rosuvastatin is more susceptible to
alkaline degradation. After 24 hours, degradation has reached 6.78% and 7.64%
for Ezetimibe and Rosuvastatin, respectively. There was no significant change
in chromatographic separation during the course of the study with the
degradation products being well separated from the parent drug peaks. The
results revealed the developed method can accurately estimate both drugs in the
presence of any alkaline degradation products.
Oxidative
Degradation
Hydrogen
peroxide was used to investigate the oxidative stability of Rosuvastatin and
Ezetimibe during the oxidative degradation studies. In cases of oxidative
stress, Ezetimibe was oxidized by 4.51% after 2 hours compared with Rosuvastatin
which was degraded by 6.99% in the same period, suggesting that Rosuvastatin
was more sensitive to oxidation. Ezetimibe degradation also rose after 24
hours, reaching 7.32%, and this degradation also increased for Rosuvastatin
reaching 8.99% after 24 hours. The chromatograms under oxidative condition
indicated complete separation of all the degradation products from the drug
peaks, and the developed RP-HPLC method was found to be specific and oxidative
stress indicative.
Table 16 Forced Degradation Study of Ezetimibe (9 µg/mL)
After 2 Hours
|
Sr. No.
|
Degradation
Condition
|
Area of
Standard
|
Area of
Degraded Sample
|
Degraded
up to (%)
|
Actual
% Degradation
|
|
1
|
Acid Degradation
|
989.60168
|
951.68182
|
96.17
|
3.83
|
|
2
|
Basic Degradation
|
989.60168
|
950.30908
|
96.03
|
3.97
|
|
3
|
H₂O₂ Degradation
|
989.60168
|
944.95886
|
95.49
|
4.51
|
Table 17 Forced Degradation Study of Rosuvastatin (18
µg/mL) After 2 Hours
|
Sr. No.
|
Degradation
Condition
|
Area of
Standard
|
Area of
Degraded Sample
|
Degraded
up to (%)
|
Actual
% Degradation
|
|
1
|
Acid Degradation
|
2121.88831
|
1977.79907
|
93.21
|
6.79
|
|
2
|
Basic Degradation
|
2121.88831
|
1976.52637
|
93.15
|
6.85
|
|
3
|
H₂O₂ Degradation
|
2121.88831
|
1973.59595
|
93.01
|
6.99
|
Table 18 Forced Degradation Study of Ezetimibe (9 µg/mL)
After 24 Hours
|
Sr. No.
|
Degradation Condition
|
Area of Standard
|
Area of Degraded Sample
|
Degraded up to (%)
|
Actual % Degradation
|
|
1
|
Acid Degradation
|
989.60168
|
919.42950
|
92.91
|
7.09
|
|
2
|
Basic
Degradation
|
989.60168
|
922.53821
|
93.22
|
6.78
|
|
3
|
H₂O₂ Degradation
|
989.60168
|
917.17700
|
92.68
|
7.32
|
Table 19 Forced Degradation Study of Rosuvastatin (18
µg/mL) After 24 Hours
|
Sr. No.
|
Degradation Condition
|
Area of Standard
|
Area of Degraded Sample
|
Degraded up to (%)
|
Actual % Degradation
|
|
1
|
Acid Degradation
|
2121.88831
|
1917.74927
|
90.38
|
9.62
|
|
2
|
Basic
Degradation
|
2121.88831
|
1959.70227
|
92.36
|
7.64
|
|
3
|
H₂O₂ Degradation
|
2121.88831
|
1931.08215
|
91.01
|
8.99
|
Photolytic
Degradation Study (UV Light)
Under
the conditions of UV light-induced stress, the drug samples have been exposed
to the UV light and subjected to photolytic degradation studies as an
assessment of the stability of the drug. Both of the drugs, Ezetimibe and
Rosuvastatin, showed less degradation with 1.65% for Ezetimibe and 1.79% for
Rosuvastatin suggesting their relative stability under photolytic conditions.
The chromatograms revealed that there was no interference from the degradation
products and there was sharpness and good separation of the drug peaks. As per
these observations, developed HPLC method proved to be good
stability-indicating HPLC method as it was able to selectively estimate
Rosuvastatin and Ezetimibe even after exposure to UV light.
Table 20 Photolytic Degradation Study of Ezetimibe and
Rosuvastatin
|
Sr. No.
|
Drug
|
Area of Standard
|
Area of Degraded Sample
|
Degraded up to (%)
|
Actual % Degradation
|
|
4
|
Ezetimibe
|
989.60168
|
973.30475
|
98.35
|
1.65
|
|
5
|
Rosuvastatin
|
2121.88831
|
2083.81470
|
98.21
|
1.79
|
CONCLUSION:
A
Stability-Indicating RP-HPLC method for the simultaneous determination of
Rosuvastatin and Ezetimibe in bulk drug and pharmaceutical dosage forms has
been developed and validated successfully which could be used for separation,
quantitative estimation of both the compounds under the study in real samples
with pretreatment method. The proposed method has good linearity, sensitivity,
accuracy, precision, specificity, ruggedness and robustness with all validation
parameters in accordance with the guideline of ICH Q2 (R2). The proposed method
was successfully employed in the assay of market tablet formulations and
satisfactory recovery and reproducible obtained. In addition, forced
degradation experiments under acidic, alkaline, oxidative and photolytic stress
conditions showed that the method is stability indicating since the peaks of
the degradation products have been separated from the analyte peak. Hence the
proposed RP-HPLC method could be applied to routine quality control, stability
and simultaneous estimation of Rosuvastatin and Ezetimibe in combined
pharmaceutical dosage form.
REFERENCES:
1.
Mukthinuthalapati MA, Bukkapatnam V,
Bandaru SP. Stability indicating liquid chromatographic method for the
simultaneous determination of rosuvastatin and ezetimibe in pharmaceutical
formulations. Adv Pharm Bull. 2014 Dec; 4(4):405-11.
2.
Varghese SJ, Ravi TK. Determination of
rosuvastatin and ezetimibe in a combined tablet dosage form using
high-performance column liquid chromatography and high-performance thin-layer
chromatography. J AOAC Int. 2010 Jul-Aug; 93(4):1222-7.
3.
Sharma S, Gaur K, Gupta R. Trends in
epidemiology of dyslipidemias in India. Indian Heart J. 2024 Mar; 76(Suppl
1):S20-S28.
4.
Phan BA, Dayspring TD, Toth PP. Ezetimibe
therapy: mechanism of action and clinical update. Vasc Health Risk Manag. 2012;
8:415-27.
5.
Bajaj T, Patel P, Giwa AO. Rosuvastatin
[Internet]. In: StatPearls. Treasure Island (FL): StatPearls Publishing; 2026
[updated 2026 Jan 31; cited 2026 Jul 16]. Available from:
https://www.ncbi.nlm.nih.gov/books/NBK539883/
6.
Sizar O, Nassereddin A, Talati R.
Ezetimibe [Internet]. In: StatPearls. Treasure Island (FL): StatPearls
Publishing; 2026 [updated 2023 Aug 28; cited 2026 Jul 16]. Available from:
https://www.ncbi.nlm.nih.gov/books/NBK532879/
7.
Davidson MH, et al. Ezetimibe
coadministered with simvastatin in patients with primary hypercholesterolemia.
J Am Coll Cardiol. 2002; 40(12):2125.
8.
Sudhop T, et al. Inhibition of intestinal
cholesterol absorption by ezetimibe in humans. Circulation. 2002;
106(15):1943-8.
9.
Sweeney ME, Johnson RR. Ezetimibe: an
update on the mechanism of action, pharmacokinetics and recent clinical trials.
Expert Opin Drug Metab Toxicol. 2007 Jun; 3(3):441-50.
10.
Kosoglou T, Statkevich P, Johnson-Levonas
AO, Paolini JF, Bergman AJ, Alton KB. Ezetimibe: a review of its metabolism,
pharmacokinetics and drug interactions. Clin Pharmacokinet. 2005; 44(5):467-94.
11.
Simard C, Turgeon J. The pharmacokinetics
of ezetimibe. Can J Clin Pharmacol. 2003 Winter; 10(Suppl A):13A-20A.
12.
LiverTox: Clinical and Research
Information on Drug-Induced Liver Injury [Internet]. Bethesda (MD): National
Institute of Diabetes and Digestive and Kidney Diseases; 2012-. Ezetimibe
[updated 2021 Dec 1; cited 2026 Jul 16]. Available from:
https://www.ncbi.nlm.nih.gov/books/NBK548095/
13.
Luvai A, Mbagaya W, Hall AS, Barth JH.
Rosuvastatin: a review of the pharmacology and clinical effectiveness in
cardiovascular disease. Clin Med Insights Cardiol. 2012; 6:17-33.
14.
Dongala T, Katari NK, Palakurthi AK,
Jonnalagadda SB. Development and validation of a generic RP-HPLC PDA method for
the simultaneous separation and quantification of active ingredients in cold
and cough medicines. Biomed Chromatogr. 2019 Nov; 33(11):e4641.
15.
Shabir GA. Development and Validation of a
Reversed-phase HPLC Method for the Determination of Hydroxybenzene in a Cream
Formulation. Indian J Pharm Sci. 2010 May; 72(3):307-11.
16.
Rani GT, Sankar DG, Kadgapathi P,
Satyanarayana B. A Validated RP-HPLC Method for Simultaneous Estimation of
Atenolol and Indapamide in Pharmaceutical Formulations. J Chem. 2011; 8:121420.
17.
Shamim A, Ansari MJ, Aodah A, Iqbal M,
Aqil M, Mirza MA, et al. QbD-Engineered Development and Validation of a RP-HPLC
Method for Simultaneous Estimation of Rutin and Ciprofloxacin HCl in Bilosomal
Nanoformulation. ACS Omega. 2023 Jun 8; 8(24):21618-21627.
18.
Draft Guidance for Industry on Analytical
Procedures and Methods Validation: Chemistry, Manufacturing, and Controls
Documentation. Fed Regist. 2000 Aug 30.
19.
Chaitali Dhale, Onkar Doke, Abhijeet B.
Survase, Uniket Gosavi, Saundarya S. Kurde. Stability Indicating Hptlc Method
Development And Validation For Simultaneous Estimation Of Rosuvastatin Calcium
And Ezetimibe In Bulk Form And Pharmaceutical Formulation. J ReAttach Ther Dev
Divers. 2023; 6(6s):737-747.
20.
International Council for Harmonisation of
Technical Requirements for Pharmaceuticals for Human Use (ICH). ICH Harmonised
Guideline: Validation of Analytical Procedures: Text and Methodology Q2(R2).
Geneva: ICH; 2023.
21.
Snorek SM, Bauer JF, Chidambaram N, et al.
PQRI Recommendations on Particle Size Analysis of Drug Substances Used in Oral
Dosage forms. J Pharm Sci. 2007; 96(6):1451-1467.
22.
Wang X, Li W, Rasmussen HT. Orthogonal
method development using hydrophilic interaction chromatography and
reversed-phase high-performance liquid chromatography for the determination of
pharmaceuticals and impurities. J Chromatogr A. 2005 Aug 12; 1083(1-2):58-62.
23.
Jawhari AH, Mohamed ZH. Developing and
Validating a Robust RP-HPLC Method for Metoclopramide and Camylofin
Simultaneous Analysis Using Response Surface Methodology. Int J Anal Chem. 2025
Oct 15; 2025:5543392.
24.
Cauchon N, Pelletier MB. Overview of ICH
Q2(R2)/Q14: Development and Validation of Analytical Procedures [Internet].
2023 [cited 2024 Jan]. Available from: https://www.casss.org/docs/default-source/consultants'-network/2023/cauchon-nina-amgen-inc.-and-pelletier-mary-beth-biogen-2023.pdf?sfvrsn=aba028c1_5
25.
Azarian S, Shaghaghi M, Dehghan G,
Sheibani N. A rapid, simple and ultrasensitive spectrofluorimetric method for
the direct detection of metformin in real samples based on a nanoquenching
approach. J Lumin. 2021; 36(3):658-667.
26.
Usha Rani N, Nagoor Ankani. New
stability-indicating RP-HPLC method development and validation for the
determination of rosuvastatin (calcium) in pharmaceutical dosage forms. Int J
Pharm Sci Drug Res. 2014; 6(4):283-290.
27.
Ramu K, Aleti P, Venisetty RK. Analytical
method development and validation of simultaneous estimation of ezetimibe and
rosuvastatin in tablet dosage form by RP-HPLC. Int J Pharm Biol Sci. 2012;
3(4).
28.
Sharma S, Sharma MC, Kohli DV, Chaturvedi
SC. Simultaneous estimation of Rosuvastatin calcium and Ezetimibe in tablet
dosage form by reverse phase high performance liquid chromatography.
Optoelectron Adv Mater Rapid Commun. 2010 Feb; 4:238-241.
29.
Kurbanoglu S, Esim O, Kose Ozkan C, Uslu
B, Savaser A, Ozkan SA. Stability-indicating liquid chromatographic method for
the simultaneous determination of rosuvastatin and ezetimibe from
pharmaceuticals and biological samples. JOTCSA. 2020; 7(3):865-874.
30.
Gupta A, Mishra P, Shah K. Simple UV
spectrophotometric determination of Rosuvastatin calcium in pure form and in
pharmaceutical formulations. E J Chem. 2009; 6:89-92.
31.
Hasumati AR, Rajput SJ, Dave JB, Patel CN.
Development and validation of two chromatographic stability-indicating methods
for determination of Rosuvastatin in pure form and pharmaceutical preparation.
Int J Chem Tech Res. 2009; 1:677-689.